beautypg.com

How to change the settings on your s8 escape ii, How to change the ramp time, How to change the epr level – ResMed Positive Airway Pressure Device S8 Escape II User Manual

Page 10

background image

6

How to Change the Settings on your S8 Escape II

You can change settings (such as Ramp Time) on the S8 Escape II by using the
keypad and LCD. When the S8 Escape II is in standby mode you can display a
series of screens (a menu) on the LCD screen in order to view and change the
settings for a particular function.

Use the menu screens to change the ramp time and language. You can also view
details about the usage and software version of your S8 Escape II.

How to change the ramp time

During ramp time, the pressure increases from a low pressure to the prescribed
treatment pressure. You can set the ramp time in 5-minute intervals, from 0
minutes (OFF) up to a maximum ramp time set by your clinician. The standby
screen displays the current ramp time in the right corner.

1. Press

or

until the desired ramp time appears.

2. Once a ramp time is selected, begin therapy by pressing

.

When ramping begins the LCD will display the word "RAMPING", with the
current pressure indicated as dashes. Each dash is approximately equivalent
to 2 cm H

2

O. For example, six dashes represents approximately 12 cm H

2

O

pressure.

How to change the EPR level

EPR can be used to help ease any discomfort you may experience when
breathing out during treatment. If your clinician has allowed access you can
change the EPR level. There are four settings: OFF; 1 (lowest EPR); 2; and 3
(highest EPR).

S8 CLASSIC

S8 CLASSIC

S8 ESCAPE II

PATIENT MENU

RAMP

10min

Menu

EPR LEVEL: 1
Change

Exit

USED HRS:

960

Exit

USAGE:

120/180

Exit

LANG:

ENGLISH

Change

Exit

SW:

SX123456789

Exit

Only displayed if
clinician enabled

338518 American Multi.book Page 6 Tuesday, February 5, 2008 2:07 PM