HEARTSTART M5066A

Patient treatment protocol options

parameter

settings

default

default description

“call EMS” voice

reminder timing

• At power on (when

the user turns on
the HeartStart)

• At power on and at

the start of the first
patient care pause

• At the start of the

first patient care
pause

• No reminder

At the start of
the first
patient care
pause

Provides a voice reminder to make
sure emergency medical services have
been called, at the start of the first
patient care pause.

shock series

1, 2, 3, 4

1

The automatic protocol pause for CPR
is activated each time a shock is
delivered.

*

During the protocol pause, the
HeartStart does not perform rhythm
analysis.

The length of the protocol pause after
a shock series is completed is
determined by the protocol pause
timer setting.

shock series interval

(minutes)

1.0, 2.0,

(infinity)

1.0

A delivered shock must occur within 1
minute of the previous shock to be
counted as part of the current shock
series.

NOTE: This parameter is only applicable
when the shock series is not configured to
the default 1 shock.

* A shock series begins when a shock is delivered after the HeartStart is turned on. A new shock series begins after a

protocol pause. If shock series is configured for 2 or more, a new shock series also begins if the time since the previous
shock exceeds the shock series interval setting.