beautypg.com

Step 3 : press the shock button – Philips M5066A User Manual

Page 22

background image

3-4

Ph
ilip

s M
e

di

ca

l Sy
st

e

m

s

HEARTSTART M5066A

STEP 3: PRESS the Shock button

As soon as the HeartStart detects that the pads are attached to the patient, it

begins analyzing the patient’s heart rhythm. It tells you that no one should be

touching the patient, and the Caution light

begins flashing as a reminder.

If a shock is needed:

The Caution light

goes from flashing to solid, the orange Shock button

starts flashing, and the defibrillator tells you to press the flashing orange button.

Before you press the button, make sure no one is touching the patient. When you

press the Shock button, the defibrillator tells you that the shock has been

delivered. Then the HeartStart tells you it is safe to touch the patient, instructs

you to begin CPR, and invites you to press the flashing blue i-button

for CPR

Coaching if desired.

If a shock is not needed:

The HeartStart tells you it is safe to touch the patient and instructs you to

perform CPR if needed. (If CPR is not needed – for example, if the patient is

moving or regaining consciousness – follow your local protocol until emergency

medical personnel arrive.) Then the HeartStart invites you to press the flashing

blue i-button

for CPR Coaching, if desired.

For CPR Coaching:

Press the flashing blue i-button

during the first 30 seconds of the patient care

pause to activate CPR Coaching.

*

(If the Infant/Child SMART Pads Cartridge is

inserted, CPR Coaching will provide coaching for infant/child CPR.) When the

pause is over, the defibrillator tells you to stop CPR, so it can analyze the patient’s

heart rhythm. The motion caused by CPR can interfere with analysis, so be sure to

stop all motion when instructed.

* The default configuration for the HeartStart provides CPR Coaching when you press the

i-button in this situation; however, the default setting can be revised by your Medical
Director using Philips software available separately. See Appendix F for more
information.