HeartStart Defibrillator specifications

The specifications provided in the following tables are nominal values. Additional
information can be found in the Technical Reference Manual for HeartStart
Automated External Defibrillators, located online at www.medical.philips.com.

Physical

Environmental

category

specifications

size

2.80” H x 7.40” D x 8.30” W (7.1cm H x 19cm D x 21cm W).

weight

Approximately 3.3 lbs (1.5 kg) with battery and pads cartridge installed.

category

specifications

temperature and

relative humidity

Operating (battery and pads cartridge installed):
32° to 122° F (0° to 50° C)
0% to 95% RH (non-condensing).

Standby (battery and pads cartridge installed):
50° to 109° F (10° to 43° C)
10% to 75% RH (non-condensing).

Storage/shipping (with battery and pads cartridge):
-4° to 140° F (-20° to 60° C) for up to 2 days
0% to 85% RH (non-condensing)

altitude

Operates at 0 to 15,000 feet; can be stored at up to 8,500 feet, in standby
mode.

shock/drop abuse

tolerance

Withstands 1 meter drop to any edge, corner, or surface.

vibration

Operating: meets EN1789 random, road ambulance.

Standby: meets EN1789 swept sine, road ambulance.