V I I I . C A R E A N D M A I N T E N A N C E

! All wheel bearings are prelubricated and sealed. They require no subsequent lubrication.
! The body shroud has been sprayed with a clear sealant coating. You can apply a light coat of car wax to help it retain

its high-gloss appearance.

! Check all electrical connections. Make sure they are tight and are not corroded. Batteries must sit flat within the battery

well, with the battery terminals facing inward, toward each other. Refer to the frame decal for the correct wiring layout.

WARNING! Even though the power chair has passed the necessary testing requirements for ingress

of liquids, you should keep electrical connections away from sources of dampness, including

direct exposure to water or bodily fluids and incontinence. Check electrical components frequently

for signs of corrosion and replace as necessary.

Daily Checks

! With the controller turned off, check the joystick. Make sure it is not bent or damaged and that it returns to the center

when you release it. Check the rubber boot around the base of the joystick for damage. Visually inspect the boot. Do
not handle or try to repair it. See your Quantum Rehab Specialist if there is a problem.

! Visually inspect the controller harnesses. Make sure that they are not frayed or cut or have any wires exposed. See

your Quantum Rehab Specialist if there is a problem with any of these harnesses.

Weekly Checks

! Disconnect and inspect the controller and the charger harnesses from the electronics tray. Look for corrosion. Contact

your Quantum Rehab Specialist if necessary.

! Ensure that all parts of the controller system are securely fastened to your power chair. Do not overtighten any screws.
! Check for proper tire inflation. There should be 35 psi (2.4 bar) in each tire. If a tire does not hold air, see a Quantum

Rehab Specialist for replacement of the tube.

! Check the brakes. This test should be carried out on a level surface with at least three feet of clearance around your

power chair.

To check the brakes:
1. Turn on the controller and turn down the speed level of your power chair.
2. After one second, check the battery condition meter. Make sure that it remains on.
3. Slowly push the joystick forward until you hear the electric brakes click. Immediately release the joystick. You must be

able to hear each electrical brake operating within a few seconds of joystick movement. Repeat this test three times,
pushing the joystick backwards, then left, and then right.

Monthly Checks

! Check that the anti-tip wheels do not rub the ground when you are operate your power chair. Adjust them as neces-

sary. See V. “Comfort Adjustments.”

! Check for extreme wear on the anti-tip wheels. Replace them as necessary.
! Check for drive tire wear. See a Quantum Rehab Specialist for repair.
! Check the rear casters for wear. Replace them as necessary.
! Check the rear forks for damage or fluttering which indicates that they may need to be adjusted or have the bearings

replaced. See a Quantum Rehab Specialist for repair.

! Keep your power chair clean and free of foreign material, such as mud, dirt, hair, food, drink, etc.

Yearly Checks

Take your power chair to a Quantum Rehab Specialist for yearly maintenance, especially if you use your power chair on a
daily basis. This helps ensure that your power chair is functioning properly and helps prevent future complications.