1) installing the range7/5, Setup procedure – Konica Minolta Non-Contact 3D Digitizer Range7 User Manual
Page 14
![background image](/manuals/156270/14/background.png)
12
This section describes the procedure for installing the RANGE7/5 properly, and completing the preparations for measurement.
Make preparations according to the following procedures:
1) Installing the RANGE7/5.
2) Set the RANGE7/5 at a position relative to the proper target, and mount the lens suitable in question as the target of measure. (No
replacement lens for RANGE5.)
3) Connect the AC adaptor to the RANGE7/5.
4) Connect the RANGE7/5 to a PC.
5) Turn ON the RANGE7/5 power switch to start warming up the RANGE7/5.
1) Installing the RANGE7/5
The RANGE7
/5 can be directly installed on a table by using the rubber soles at the bottom. Also, it can be installed with the tripod set or
measuring stand (under optional accessories). When the tripod set or measuring stand are used, the RANGE7
/5 orientation can be freely
changed (up/down and right/left). This feature is useful for measurement.
Memo For installation procedure details, please refer to the description on “Installation”.
2) Setting the RANGE7/5’s position relative to the measuring target, and exchanging the lens
(RANGE7 only)
Scan Position Adjustment
The location and posture of the work and RANGE7/5 can be adjusted so that the two are appropriately positioned for
scanning. These adjustments are made by activating the monitoring feature and watching the work in the monitor window.
Click the [Monitor] button to check the work on the monitor
window.
Distance to work and work size
(Unit: mm)
Lens
TELE lens
WIDE lens
Distance
450
800
450
800
XЧY
79Ч99
141Ч176
150Ч188
267Ч334
Z
54
97
109
194
Memo For RANGE5, please refer to the WIDE lens data.
Distance to work and work size
in MultiFocus mode
(Unit: mm)
Lens
TELE lens
WIDE lens
Distance
462
781
475
766
XЧY
81Ч102
138Ч172
159Ч199
256Ч320
Z
54
97
109
194
Memo For RANGE5, please refer to the WIDE lens data.
Distance to work
450mm to 800mm
Receiver lens
RANGE7/5
Work
Y
X
Z
• The origin is set at the sensor position which is
approximately 120 mm behind the center of the lens
surface. Therefore, the Z dimension has an offset of
around 120 mm applied.
RANGE7: TELE/WIDE replaceable
RANGE5: No replacement lens
RANGE VIEWER monitor window