MAINTENANCE

VOYAGER 420 - 550 – 800 (rev 3)

16

7 MAINTENANCE

ALWAYS CARRY OUT THE

DAILY CHECKLIST BEFORE
EACH LIFT USE.

Alterations made to the VOYAGER 420,
550 or 800 by someone other than a
certified technician may cause serious
injury and voids warranty.

The VOYAGER 420, 550, 800 and
accessories must be inspected
ANNUALLY by a certified technician in
addition to the daily and other periodic
visual checks done by the user
specified in this section.

Preventive maintenance specified in this
manual can prevent accidents and
reduce repair costs.

Note all service or repairs to the
VOYAGER 420, 550, 800 or its
accessories in the log book at the end of
this manual. Have the document signed
by the certified technician.

7.1 Daily Check List

THE FOLLOWING PROCEDURES MUST
BE FOLLOWED BEFORE EACH USE.

• Has the battery been charged? Park the lift

on the charger whenever the lift is not in use.

• Inspect the lift for any damage. If the lift

casing does not look properly aligned, or
there are any cracks or other damage on the
lift, or any parts are missing - DO NOT USE
IT. Contact your local representative to
have the lift serviced.

• Inspect the visible strap for any signs of

wear, frays, loose threads or other damage.
If there is any evidence of damage - DO
NOT USE IT. Contact your local
representative to have the lift serviced.

• Inspect the sling for tears, frayed straps or

loose stitching. If the sling has any of the
above damage –DO NOT USE IT. Contact
your local representative to have the sling
replaced or repaired.

• Inspect the carry bar for any signs of

cracking or damage.

• Ensure the ring and cotter pins that attach

the carry bar to the strap are secure.

7.2 Inspection and Cleaning

Clean the Voyager with a soft dry cloth, or a soft
cloth lightly moistened with a mild detergent
solution. Do not use any type of solvent that may
damage the finish.

Do not immerse lift in water.

To ensure a better rolling surface for the trolley
wheels, clean the inside of the track every 4
months. To do so, insert a damp cloth in the
opening and slide it from one end of the track to
the other.

Always reinstall the rail end
stopper (if they have been
removed) after servicing.

7.3 Strap Inspection

This equipment is built with a strap that can
withstand a load of 1364 kg (3000 lbs.)

Nonetheless, if this strap is damaged or shows
signs of wear, the acceptable load on the strap
before rupture can drop rapidly and present a
danger for the patient or caregiver.

BHM Medical recommends thoroughly inspecting
the strap every 2 months as follows:
1. Completely

unwind

the

strap.

2.

Look for any signs of wear:











If there is any signs of wear as
indicated here or other visual
defects, strap should be
changed immediately. By
continuing to use the lift without
changing the strap, caregiver
and patient safety is greatly
compromised.

In any case, the manufacturer

recommends changing the strap at least
every two years. By continuing to use the
lift without changing the strap, caregiver
and patient safety is greatly

Loose threads in stitched area.

Noticeable discoloration by having a
Lighter color strap than the double

Side wear.

Middle wear.

thickness stitched area.