Sony LMD-2140MD User Manual
Page 3
3
Important safeguards/notices for use in the
medical environments
1. All the equipment connected to this unit shall be
certified according to Standard IEC60601-1,
IEC60950-1, IEC60065 or other IEC/ISO Standards
applicable to the equipment.
2. When this unit is used together with other equipment
in the patient area*, the equipment shall be either
powered by an isolation transformer or connected via
an additional protective earth terminal to system
ground unless it is certified according to Standard
IEC60601-1 and IEC60601-1-1.
3. The leakage current could increase when connected
to other equipment.
4. The operator should take precautions to avoid
touching the rear panel input and output circuitry and
the patient at the same time.
5. Model LMD-2140MD is a monitor intended for use
in a medical environment to display pictures from
cameras or other systems. These models are
nonpatient care equipment with respect to the
requirement of leakage current.
Caution
When you dispose of the unit or accessories, you must
obey the law in the relative area or country and the
regulation in the relative hospital.
Symbols on the unit
R1.5m
* Patient Area
Symbol
Location
This symbol indicates
Bottom
Main power switch.
Press to turn the monitor
on.
Bottom
Main power switch.
Press to turn the monitor
off.
Bottom
The equipotential
terminal which brings the
various parts of a system
to the same potential.
Bottom
Functional earth terminal
8
Front
Brightness control
6
Front
Contrast control
-
Front
Key inhibit
The setting are locked so
that they cannot be
changed.
Warning on power connection
Use a proper power cord for your local power supply.
The United States
Canada
Continental Europe
Plug type
HOSPITAL GRADE
E41395
E41395-A
HOSPITAL GRADE
LL33182
LL76662
LP-34A
Female end
LS-60
Cord type
H05VV-F
Rated Voltage & Current
10A/125V
10A/125V
10A/250V
Safety approval
UL
CSA
VDE