SECTION 2—EMI INFORMATION

Part No. 1118395

19

Zoom

300 and Zoom

400

ƽ WARNING

Powered Wheelchair Electromagnetic Interference (EMI)
Because EM energy rapidly becomes more intense as one moves closer to the

transmitting antenna (source), the EM fields from hand-held radio wave sources

(transceivers) are of special concern. It is possible to unintentionally bring high

levels of EM energy very close to the powered wheelchair's control system while

using these devices. This can affect powered wheelchair movement and braking.

Therefore, the warnings listed below are recommended to prevent possible

interference with the control system of the powered wheelchair.
Electromagnetic interference (EMI) from sources such as radio and TV stations,

amateur radio (HAM) transmitters, two-way radios, and cellular phones can affect

powered wheelchairs and motorized scooters.
FOLLOWING THE WARNINGS LISTED BELOW SHOULD REDUCE THE

CHANCE OF UNINTENDED BRAKE RELEASE OR POWERED WHEELCHAIR

MOVEMENT WHICH COULD RESULT IN SERIOUS INJURY.
1) Do not operate hand-held transceivers (transmitters receivers), such as citizens

band (CB) radios, or turn ON personal communication devices, such as cellular

phones, while the powered wheelchair is turned ON;

2) Be aware of nearby transmitters, such as radio or TV stations, and try to avoid

coming close to them;

3) If unintended movement or brake release occurs, turn the powered wheelchair

OFF as soon as it is safe;

4) Be aware that adding accessories or components, or modifying the powered

wheelchair, may make it more susceptible to EMI (NOTE: There is no easy way

to evaluate their effect on the overall immunity of the powered wheelchair); and

5) Report all incidents of unintended movement or brake release to the powered

wheelchair manufacturer, and note whether there is a source of EMI nearby.

Important Information
1) 20 volts per meter (V/m) is a generally achievable and useful immunity level

against EMI (as of May 1994) (the higher the level, the greater the protection);

2) This device has been tested to a radiated immunity level of 20 V/m.
Modification of any kind to the electronics of this wheelchair as manufactured by

Invacare may adversely affect the RFI immunity levels.