7 MAINTENANCE AND TROUBLESHOOTING

Invacare®MA60 Series

34

Part No 1148137

7.2

Troubleshooting

PROBLEM

CAUSE

SOLUTION

Mattress not inflating
Not alternating properly

Mattress hose disconnected

Connect hose connectors and lock them in place

Air hose kinked or split

Unkink hose or replace split hose

Major leak in the air cushions or overlay pad

Replace leaking air cushions or overlay pad

Kinked or split manifold

Unkink manifold or replace split manifold

Has power and fuse is good, control unit does not
come on

Send control unit back to the factory for repair

Not alternating, solenoid malfunction

Send control unit back to the factory for repair

No air, pump malfunction

Send unit for repair

No power

Control unit off

Check power source and turn on unit

Power cord disconnected

Connect power cord to the power source

No power in the power source

Check power source has power and turn it On

Power outage

Wait till the power source has power

Blown fuse

Replace blown fuse with an equivalent fuse