6 OPERATION
28
9.
Verify that the patient has not bottomed out by performing the following steps:
A. Ensure that the patient is lying flat on his/her back in the middle of the mattress.
B.
Place four fingers between the air cushions directly underneath the sacral region of the patient’s body.
C. Ensure that there is 3 to 4-finger width clearance between the bottom of the patient and the mattress surface.
D. Adjust the comfort setting, if needed.
E.
Wait five minutes for the mattress pressure to stabilize.
F.
Repeat STEPS A-E until patient has not bottomed out and patient comfort is achieved.
10. If the patient feels that the bed is too soft/hard, press the Soft or Firm button to adjust the comfort settings.
11. Use a regular pillow to help support and stabilize the patient's head.
6.9
Transferring Patient From/To a Gurney
1.
Engage the wheel locks of the bed. Refer to the owner’s manual provided with the bed.
2.
Engage the wheel locks of the gurney. Refer to the owner’s manual provided with the gurney.
3.
Press the Max Inflate button to achieve maximum mattress pressure.
4.
Raise or lower the bed to match the gurney height. Refer to the owner’s manual provided with the bed.
5.
When the mattress has reached maximum firmness, perform one of the following:
•
Bed to Gurney Transfer - Slide the patient onto the gurney.
•
Gurney to Bed Transfer - Slide the patient onto the bed.
ƽ
WARNING
ALWAYS engage the wheel locks of the bed and the wheel locks of the gurney before transferring the patient between the
bed and the gurney.