Troubleshooting – Invacare IRC5LXO2 User Manual
Page 20
20
PART
ONE
TROUBLESHOOTING
1. Internal repairs required. 1. Call Service Provider.
GREEN LIGHT with
YELLOW Light
Flashing.
Alarm:
RAPID
Beep....Beep...
Beep...Beep
LOW FLOW ALARM
1a. Kinked or blocked
tubing, cannula or
humidifier.
1b. Flowmeter set at
0.5 L/min. or less.
1a. Inspect for kinks or
blockages. Correct,
clean or replace item.
Once corrected, turn
power OFF for 60
seconds and then turn
power back ON.
1b. Reset flowmeter to
prescribed flowrate. If
prescribed flowrate is
below 0.5 L/min., use
the Pediatric
Flowmeter IRCPF16.
SYMPTOM:
YELLOW or RED
LIGHT Illuminated.
Alarm:
CONTINUOUS
On red light only.
PROBABLE CAUSE:
1. Low oxygen purity.*
2. Kinked or blocked
tubing, cannula or
humidifier.*
3. Flowmeter set at
1/2 L/min. or less.*
4. Unit overheating
due to blocked air
intake.
5. Insufficient power at
outlet.
6. Internal repairs required.
SOLUTION:
1. Clean or Replace filters.
2. Inspect for kinks or
blockages. Correct,
clean or replace item.
Once corrected, turn
power OFF for 60
seconds and then turn
power back ON.
3. Check flowmeter is set
to 1.0 L/min. or more.
Refer to Recommended
End-User Guidelines in
the Operating
Instructions of this
manual.
4.a.Remove and clean
cabinet filters.
4b. Move concentrator at
least 3- inches from
walls, draperies, and
furniture.
5. Do not use extension
cords. Move to another
electrical outlet or
circuit.
6. Call Service Provider.
* Only applies to
SensO
2
models.