Concentrator maintenance – Invacare IRC5LXO2 User Manual
Page 18
18
PART
ONE
Cabinet Filter
(located on both sides
of concentrator)
NOTE: There are two (2) cabinet filters. One (1) is located on each side of
the cabinet (FIGURE 1).
1. Remove each filter and CLEAN AT LEAST ONCE A WEEK
depending on environmental conditions (FIGURE 2).
CONCENTRATOR MAINTENANCE
MAINTENANCE
FIGURE 1 - CABINET FILTER LOCATION
NOTE: The Invacare Concentrators are specifically designed to minimize
routine preventive maintenance at intervals of once per year. In places with high
dust or soot levels, maintenance may need to be performed more often. The
following must be performed at a minimum of one (1) year in service to assure
years of additional reliability. Only authorized or factory-trained personnel should
perform preventive maintenance on the concentrator. Power should be discon-
nected before beginning preventive maintenance on the Invacare concentrator.
CLEANING THE CABINET FILTERS
NOTE: Environmental conditions that may require more frequent cleaning
of the filters include but are not limited to: high dust, smoking, air pollutants, etc.
2. Clean the cabinet filters with a vacuum cleaner or wash in
warm soapy water and rinse thoroughly (FIGURE 2).
3. Dry the filters thoroughly before reinstallation.
CAUTION
DO NOT operate the concentrator without the filters installed.
FIGURE 2 - CLEANING THE CABINET FILTERS