Liquid level detector, Liquid level detector -8 – Gilson 215 User Manual
Page 79

5-7
Troubleshooting
5
Liquid Level Detector
Liquid Level Detector
Liquid level detector not detecting liquid level
• Ensure that the level sensing cables are plugged
in.
• Check sensitivity setting in the 215 Setup Utility
• Check if liquid is detectable. Liquid level detection
works only if there is electrical conductivity in
your liquid. Liquid level detecting will not work
with most non-polar liquids. For intermediate
polarity liquids and polar liquids, check the
sensitivity setting in the 215 Setup Utility.
• Call Gilson if this is caused by faulty circuitry.
Liquid level detector falsely detecting liquid
level
• Ensure that probes are installed correctly.
• Check sensitivity setting in the 215 Setup Utility
• Call Gilson if this is caused by faulty circuitry.