Appendix – Gilson 215 User Manual
Page 102

C-4
Appendix
C
GSIOC Commands
9
B
Clears the contact event FIFO and resets the contact input
event timer.
A
I
Reads home phases for X and Y motors. Returns “xxx/yyy”
where:
xxx - X motor phase.
yyy - Y motor phase.
Bf,d
B
Causes 215 to beep:
f - Frequency in Hz; default is 2400.
d - Duration of beep in tenths of seconds; default is 1.
cv=ss..ss
B
Sets user-definable character:
v
- Address (0 to 7) of user-defined character.
ss..ss - Characters, in hexadecimal ASCII code, for
rows 1 through 7. See example below.
For example, the following command sets the pattern
shown below to user-defined character address 0:
c0=1F111B151B111
[Row]
[Pattern]
[Hex Code]
1
* * * * *
1F
2
* *
11
3
* * * *
1B
4
* * *
15
5
* * * *
1B
6
* *
11
7
* * * * *
1F
e
I
Reads the current error number. Returns “nnn” which
identifies the error number; see page
errors. If no error has occurred, returns 0.
e
B
Clears error number.
Note: Whenever possible, it is recommended that the
immediate “$” command be used to clear errors.