Error messages, Error messages -2, Troubleshooting – Gilson 215 User Manual
Page 74

5-2
Troubleshooting
5
Error Messages
Error Description
Solution
15
NV-RAM checksum is invalid
• Send the buffered ~9 GSIOC command to
reset the NV-RAM and initialize to the
default value. Run 215SETUP.EXE or the
215 Setup Utility program
• Replace the main board
16
X scale factor is invalid
Contact the Gilson Customer Service Department.
17
Y scale factor is invalid
Contact the Gilson Customer Service Department.
20
X motor position error
Turn power off then on to the liquid handler.
21
Y motor position error
Turn power off then on to the liquid handler.
22
Z motor position error
Turn power off then on to the liquid handler.
24
X target less than minimum X
Send the immediate Q command using the
GSIOC Utility Program to read the travel range.
Correct the error in the program controlling the
liquid handler.
25
X target more than maximum X Send the immediate Q command using the
GSIOC Utility Program to read the travel range.
Correct the error in the program controlling the
liquid handler.
26
Y target less than minimum Y
Send the immediate Q command using the
GSIOC Utility Program to read the travel range.
Correct the error in the program controlling the
liquid handler.
27
Y target more than maximum Y Send the immediate Q command using the
GSIOC Utility Program to read the travel range.
Correct the error in the program controlling the
liquid handler.
28
Z target less than minimum Z
Send the immediate Q command using the
GSIOC Utility Program to read the travel range.
Correct the error in the program controlling the
liquid handler.
Error Messages