beautypg.com

4 safety guidelines 11, 4 safety guidelines, 4safety guidelines – Fluke 9107 User Manual

Page 17

background image

4

Safety Guidelines

• Operate the instrument in room temperatures between 5–50°C

(41–122°F). Allow sufficient air circulation by leaving at least 6 inches of
space between the instrument and nearby objects. DO NOT place under a
cabinet or other structure. Allow for overhead clearance.

• The dry-well is a precision instrument. Although it has been designed for

optimum durability and trouble free operation, it must be handled with
care. Always carry the unit in an upright position to prevent the probe
sleeves from dropping out. The convenient fold-up handle allows one
hand carrying. The instrument should not be operated in excessively wet,
oily, dusty, or dirty environments. It is important to keep the well of the
instrument clean and clear of any foreign matter. DO NOT operate near
flammable materials.

DO NOT use fluids to clean out the well.
• The instrument can generate extreme temperatures. Precautions must be

taken to prevent personal injury or damage to objects. Probes may be ex-
tremely hot or cold when removed from the instrument. Cautiously handle
probes to prevent personal injury. Always use the special sleeve tongs that
are supplied with the calibrator to remove the sleeve. Carefully place
probes on a heat/cold resistant surface or rack until they are at room tem-
perature. Never place any objects other than the special probe sleeves sup-
plied with the calibrator into the well.

• Use only a grounded AC mains supply of the appropriate voltage to

power the instrument. The Model 9105 dry-well requires 3 amps SB at
115 VAC (±10%), 50/60 Hz, 1.6 amps T at 230 VAC (±10%). The Model
9107 dry-well requires 4 amps SB at 115 VAC (±10%), 50/60 Hz, 3.15
amps T at 230 VAC (±10%).

• Before initial use, after transport, and anytime the dry-well has not been

energized for more than 10 days, the instrument needs to be energized for
a “dry-out” period of 1-2 hours before it can be assumed to meet all of the
safety requirements of the IEC 1010-1.

• The instrument is equipped with operator accessible fuses. If a fuse

blows, it may be due to a power surge or failure of a component. Replace
the fuse once. If the fuse blows a second time, it is likely caused by fail-
ure of a component part. If this occurs, contact Hart Scientific Customer
Service. Always replace the fuse with one of the same rating, voltage, and
type. Never replace the fuse with one of a higher current rating.

• If a mains supply power fluctuation occurs, immediately turn off the in-

strument. Power bumps from brown-outs and black-outs could damage

9105/9107

11

4 Safety Guidelines

This manual is related to the following products: