2 cautions 4, 2 cautions – Fluke 9107 User Manual
Page 11

• DO NOT REMOVE INSERTS AT HIGH TEMPERATURES. Inserts
will be the same temperature as the display temperature. Use extreme care
when removing hot inserts.
• DO NOT operate this unit without a properly grounded, properly polar-
ized power cord.
• DO NOT connect this unit to a non-grounded, non-polarized outlet.
• HIGH VOLTAGE is used in the operation of this equipment. SEVERE
INJURY OR DEATH may result if personnel fail to observe safety pre-
cautions. Before working inside the equipment, turn power off and dis-
connect power cord.
• Always replace the fuse with one of the same rating, voltage, and type.
• Overhead clearance is required. DO NOT place unit under a cabinet or
other structure.
• DO NOT use this unit for any application other than calibration work.
• DO NOT use this unit in environments other than those listed in the
user’s manual.
• DO NOT turn the unit upside down with the inserts in place; the inserts
will fall out of the unit.
• DO NOT operate near flammable materials.
• Use of this instrument at HIGH TEMPERATURES for extended periods
of time requires caution.
• Completely unattended high temperature operation is not recom-
mended for safety reasons.
• Before initial use, after transport and anytime the dry-well has not been
energized for more than 10 days, the calibrator must be energized for a
dry-out period of 1 to 2 hours before it can be assumed to meet all of the
safety requirements of the IEC1010-1.
• Follow all safety guidelines listed in the user’s manual.
• CALIBRATION EQUIPMENT should only be used by TRAINED PER-
SONNEL.
1.2.2
Cautions
To avoid possible damage to the instrument, follow these guidelines.
• Components and heater lifetime can be shortened by continuous high
temperature operation.
4
Hart Scientific
1 Before You Start