Detailed specifications – Fluke Thermal Imagers Ti10 User Manual
Page 23
Thermal
Imagers
Detailed
Specifications
17
Detailed Specifications
Temperature Measurements
Temperature Range (not calibrated below -10
°
C)
TiR and TiR1 ................................................. -20
°
C to +100
°
C
Ti10 ............................................................... -20
°
C to +250
°
C
Ti25 ............................................................... -20
°
C to +350
°
C
Accuracy
TiR and Ti10 ..................................................
±
5
°
C or 5 % (whichever is greater)
TiR1 and Ti25 ................................................
±
2
°
C or 2 % (whichever is greater)
Measurement Modes ......................................... Smooth Auto-Scaling and Manual
Scaling
On-screen Emissivity Correction ........................ (Ti25 and TiR1 Only)
Imaging Performance
Field of View ...................................................... 23
°
x 17
°
Spatial Resolution (IFOV) .................................. 2.5 mRad
Minimum Focus Distance
Thermal Lens................................................. 15 cm (approx. 6 in)
Visible (Visual) Light Lens.............................. 46 cm (approx. 18 in)
Focus ................................................................ Manual
Image Frequency............................................... 9 Hz refresh rate
Detector Type .................................................... 160 X 120 Focal Plane Array,
uncooled microbolometer
Infrared Lens Type............................................. 20 mm EFL, F/0.8 lens
Thermal Sensitivity (NETD)
Ti10 ...............................................................
≤
0.2
°
C at 30
°
C (200 mK)
Ti25 ...............................................................
≤
0.1
°
C at 30
°
C (100 mK)
TiR.................................................................
≤
0.1
°
C at 30
°
C (100 mK)
TiR1...............................................................
≤
0.07
°
C at 30
°
C (70 mK)
Infrared Spectral Band ....................................... 7.5
μ
m to 14
μ
m
Visual Camera ................................................... 640 x 480 resolution
Image Presentation
Palettes
TiR and Ti10 .................................................. Ironbow, Blue-Red, High Contrast,
Grey
TiR1 and Ti25 ................................................ Ironbow, Blue-Red, High Contrast
Amber, Hot Metal, Grey
Level and Span
Smooth Auto-Scaling and Manual Scaling of level and span
Minimum Span (in manual mode)