Assembly instructions, Disassembly instructions – Precision Medical PreSet Continuous Vacuum Regulator 0-200mmHg User Manual
Page 8

7
CONTINUOUS & CONTINUOUS INTERMITTENT
ASSEMBLY INSTRUCTIONS
1. To assemble, perform the “DISASSEMBLY INSTRUCTIONS” in reverse order.
Note: • Ensure the tab on the Regulator Module is inserted into the internal
slot on the Selector Module at the 9 o’clock position.
• Ensure the tab on the Control Knob Assembly at the 12 o’clock
position is aligned with the outer circumference slot on the
Selector Module at the 12 o’clock position.
2. Lubricate all O-rings and cavities with Vacuum grease (part# 1775) supplied in
the Preset Vacuum Regulator Repair Kit.
3. Repeat steps 1 through 3 of “OPERATING INSTRUCTIONS”.
4. Prior to returning Preset Vacuum Regulator to service verify preset vacuum
level with gauge reading.
DISASSEMBLY INSTRUCTIONS
for PS3300 & PS3400 Series
(Reference “PARTS DESCRIPTION”)
1. Loosen the Set Screw (Item # 11) in Selector Knob.
2. Pull the Regulator Module/Control Knob Assembly (Item # 12) away from
case. (The Regulator Module (Item # 10) is threaded onto the Control Knob
Assembly. Note: These components must always be kept together as a
matched set.)
3. Remove the screws (Item # 2) from the back of the product.
4. Remove the Rear Case (Item # 14) by pulling away from product.
5. Remove screws (Item# 2) from the top of the Timing Module.
6. Remove the Timing Module (Item# 13) by pulling away from the Housing
Assembly (Item# 1).
7. Separate the Case Assembly (Item# 9) by pulling it away from the Housing
Assembly (Item# 1).
8. Remove the Selector Assembly (Item# 6) by pulling it away from the Housing
Assembly (Item# 1).
9. Remove Gauge by pulling vertically out of socket (Item# 3).
DISASSEMBLY INSTRUCTIONS
for PS3000, PS3100, PS3500
& PS3600 Series
(Reference “PARTS DESCRIPTION”)
1. Loosen the Set Screw (Item# 11) in Selector Knob.
2.
Pull the Regulator Module/Control Knob Assembly (Item # 12) away from case.
(The Regulator Module (Item # 10) is threaded onto the Control Knob Assembly.
Note
: These components must always be kept together as a matched set.)
3. Remove the screws (Item# 2) from the back of the Regulator.
4. Separate the Case Assembly (Item# 9) by pulling it away from the Housing
Assembly (Item# 1).
5. Remove the Selector Assembly (Item# 6) by pulling it away from the Housing
Assembly (Item# 1).
6. Remove Gauge by pulling vertically out of socket (Item# 3).