Cleaning / decontamination, Maintenance, Returns – Precision Medical PreSet Continuous Vacuum Regulator 0-200mmHg User Manual
Page 10

9
CONTINUOUS & CONTINUOUS INTERMITTENT
CLEANING / DECONTAMINATION
(As needed)
1. Attach a working Vacuum Regulator with a continuous regulated
mode to a minimum vacuum source of 15 inHg.
2. Mix cold disinfection/sterilization solution according to its
manufacturer’s directions.
3. Connect tubing as shown in Cleaning Illustration on previous page.
4. Turn the working Vacuum Regulator on to a continuous regulated mode.
5. Adjust the vacuum to a minimum of 120 mmHg.
6. Set the “Preset” Vacuum Regulator to be cleaned to the “REG.”
mode, and set at 100 mmHg.
7. Allow cold disinfection/sterilization solution to pass through Preset
Vacuum Regulator and collect in Suction Canister. Procedure should
continue for time recommended by the manufacturer of the cold
disinfection/sterilization solution for the desired level of disinfection or
sterilization.
8. Turn the Preset Vacuum Regulator to be cleaned to the “LINE” mode
(if applicable)
.
9. Allow remaining cold disinfection/sterilization solution to pass through
and collect in Suction Canister.
10. Set working Vacuum Regulator to its maximum vacuum setting.
11. Thoroughly dry the internal components by drawing maximum
vacuum through the cleaned Regulator for at least 30 seconds in
both “REG.” and “LINE” modes (if applicable).
NOTE: If it is not possible to pass cold disinfection/sterilization solution
through the Regulator, then the passageways are totally blocked
and DISASSEMBLY of the Regulator is required. Ensure to follow
the Biohazard protocol at your facility.
MAINTENANCE
Before each use, visually inspect Preset Vacuum Regulator for any sign
of damage. DO NOT USE if damaged.
RETURNS
Returned products require a Returned Goods Authorization (RGA) number,
contact Precision Medical, Inc. All returns must be packaged in sealed
containers to prevent damage. Precision Medical, Inc. will not be responsible
for goods damaged in transit. Refer to Precision Medical, Inc. Return Policy
available on the Internet, www.precisionmedical.com.
Note: Federal Law prohibits shipping of BioHazard materials Via U.S. mail.
Manuals available on our Website; www.precisionmedical.com