3
CONTINUOUS & CONTINUOUS INTERMITTENT
SPECIFICATIONS
CALIBRATED RANGE
P
S3000:
0 - 200 mmHg
PS3000HV: 0 - 300 mmHg
PS3100:
0 - 200 mmHg
PS3100HV: 0 - 300 mmHg
PS3300:
0 - 200 mmHg
PS3300HV: 0 - 300 mmHg
PS3400:
0 - 150 mmHg
PS3500:
0 - 150 mmHg
PS3600:
0 - 760 mmHg
GAUGE ACCURACY
Analog Gauge
±
5% of MAX calibrated range
Dual Analog/Digital Gauge
Digital Display
± 1% of Full Scale
Analog Gauge
± 5% of MAX calibrated range within ref. Indicator
INLET/OUTLET CONNECTIONS 1/8 NPT Female
MODES OF OPERATION
PS3000 Series
PS3500 Series
OFF
- No Vacuum
REG. - (Regulated) Provides an adjustable, continuous vacuum level
PS3100 Series
PS3600Series
OFF
- No Vacuum
REG. - (Regulated) Provides an adjustable, continuous vacuum level
LINE - Provides maximum, continuous vacuum from the vacuum source
PS3300 Series
PS3400 series
OFF - No Vacuum
REG. - (Regulated) Provides an adjustable, continuous vacuum level
INT. - Intermittent) Provides an adjustable vacuum level that cycles
between ON and OFF
FLOW RATES
PS3000, PS3100 & PS3300
REG MODE @ 200 mmHg
48 l/min
LINE MODE
55 l/min
INT. MODE (PS3300 Only)
10 l/min
PS3000HV, PS3100HV & PS3300HV
REG MODE @ 300 mmHg
50 l/min
LINE MODE
55 l/min
INT. MODE (PS3300HV Only)
10 l/min
PS3300 & PS3400
REG MODE @ 150 mmHg
50 l/min
INT. Mode (PS3300 only)
10 l/min
INT. MODE (PS3400 only)
3 l/min
PS3600
REG MODE @ 635 mmHg
71 l/min
LINE MODE
82 l/min
FLOW RATES ARE
OBTAINED WITH A VACUUM SOURCE OF 21” Hg
INTERMITTENT CYCLE TIME:
Factory set at sixteen (16) seconds ON
and eight (8) seconds OFF (Reference only)
Operating Environmental Limits:
0°F to 122°F (-18°C to 50°C)
Recommended Environmental Operating Limits: 55°F to 85°F (13°C to 29°C)
Storage Environmental Limits
Temperature Range:
-4°F to 140°F (-20°C to 60°C)
Humidity:
Max 95% Noncondensing
Battery:
3 Volt Lithium,
1
/
2
AA (Digital Vacuum Gauge Models ONLY)
Specifications are subject to change without prior notice.