Inova Labs Activox DUO2 User Manual
Page 33

EMC
INFORMATION
continued
The model
130-001 is intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled. The customer or the user
of the model
130-001 can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the model
130-001
as recommended below, according to the maximum output power of the
communications equipment.
For transmitters rated at a maximum output power not listed above, the recommended
separation distance (d) in meters (m) can be estimated using the equation applicable
to the frequency of the transmitter, where p is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
Recommended Separation Distances Between Portable and Mobile RF
Communications Equipment and The Activox DUO2 Stationary Base Model
130-001.
Separation Distance According to the Frequency of Transmitter (m)
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.74
1
1.2
1.2
2.3
10
3.8
3.8
7.4
100
12
12
23
Rated Maximum
Output Power of
Transmitter (W)
150kHz to 80MHz
80MHz to 800MHz 800MHz to 2.5GHz
d=[ ] √P
d=[ ] √P
d=[ ] √P
3.5
V1
3.5
E1
7
E1
Emissions Test
Compliance
Electromagnetic Environment - Guidance
The model
130-001 use RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference to nearby electronic equipment.
RF Emissions, CISPR 11
RF Emissions, CISPR 11
Harmonic Emissions,
IEC 61000-3-2
Voltage Fluctuations/Flicker
Emissions, IEC 61000-3-3
Group 1
Class B
N/A
N/A
The model
130-001 are suitable for use
in all establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network
that supplies buildings used for domestic
purposes.