Inova Labs Activox DUO2 User Manual
Page 17

15
LIGHT INDICATORS/ALARMS
NORMAL OPERATION INDICATORS
Normal
Operation
Light Indication
Audible
Alarm
Action
Stationary
base is
plugged into a
power source.
All LEDs flash briefly.
Brief
audible
tone from
unit.
None. Stationary
base is functioning
normally.
Stationary
base is in the
“On” position
and running.
Power LED illuminates solid
green. Battery Charge LED
may illuminate or flash if
External Battery is connected.
None
Allow for a warm-up
period of approximately
2 minutes before using.
Stationary
base is
charging
the External
Battery.
The Battery Charge LED
flashes green faster as the
External Battery approaches a
full charge.
None
None. Stationary
base is functioning
normally.
Stationary
base has fully
charged the
External
Battery.
The Battery Charge LED
illuminates solid green.
None
None. Stationary
concentrator
is functioning normally.
When POC is
first connected
to the
stationary
base.
All LEDs flash once and
Pulse Setting LEDs remain
illuminated. The INT LED will
illuminate until the internal
battery is fully charged.
Brief
audible
tone from
POC.
None. POC is properly
connected to the stationary
base.