16

ALARM/LIGHT

INDICATORS

continued

ALARM INDICATORS

NORMAL OPERATION INDICATORS

Problem

Light Indication

Audible

Alarm

Action

No oxygen

flow or low

oxygen

detected by
the system.

*

Power LED is illuminated and

the Alarm LED illuminates

solid red.

5 second

audible

tone every

minute.

Remove the nasal cannula

hose from the Cannula Nozzle

Fitting:

If the alarm stops, you

have a “no oxygen flow”

alarm. Check the hose for

kinks or blockage. If found,

remove the kink or blockage

and re-connect the cannula

hose. If the alarm does not

return, continue use. If no

kink or blockage is found,

switch to the POC or another

source of oxygen and contact

your equipment provider.

If the alarm remains, you

have a “low oxygen purity”

alarm. Turn the system off.

Switch to the POC or another

source of oxygen and contact

your equipment provider.

High

pressure

detected by

the system.

The red Alarm LED light

flashes on and off every

.5 seconds

Audible

tone

alternating

on for

2 seconds

and then

off for 2

seconds.

Turn the system off.

Switch to the POC or another

source of oxygen and contact

your equipment provider.

Normal

Operation

Light Indication

Audible

Alarm

Action

Stationary

base is on and

connected to

the POC.

*

Power LED illuminates solid

green. Battery Charge LED

may illuminate or flash if

External Battery is connected.

None

None. The stationary base

is properly connected to

the POC.

*If the POC being connected to the stationary base has a completely depleted internal battery, the LEDs
on the stationary base will briefly turn off and an approximate two second audible tone will sound before
the normal operation indicators initiate.