Hardy Bench Scales User Manual

Page 13

background image

Hardy Bench Scales

3-6

Make sure the cable is not pinched and is clear
of the feet. The Bench Scale requires a recom-
mended DC or AC excitation voltage of 5 volts
with a maximum of 15 volts. Wire the Scale cable
to an instrument or indicator following the color
code label found on the cable.

For HI BS200 and HI BS300 Scale Models:

+ Excitation

Green

- Excitation

Black

+ Sense

Blue

- Sense

Brown

+ Signal

Red

- Signal

White

Shield

Bare

For HI BS400 Scale Models with C2 Electronic
Calibration:

+ Excitation

Red

- Excitation

Black

+ Signal

Green

- Signal

White

+ C2

Gray

- C2

Violet

Shield

Yellow or Bare

WARNING

USE CAUTION WHEN WIRING. MAKE SURE
TO DISCONNECT ALL POWER TO THE
SCALE. FAILURE TO REMOVE POWER CAN
RESULT IN SCALE OR INSTRUMENT DAM-
AGE, DEGRADATION OF PERFORMANCE OR
PERSONAL INJURY

NOTE:

Hardy Process Solutions provides drawings you
may download on the Hardy Bench Scales page
under the Docs & Programs tab. Drawings are
provided for each of the Hardy Bench Scale
models in .dxf, dwg, .pdf, and 3D formats. These
drawings include the wiring configurations.