Iris Sample Processing Statspin® MP Multipurpose Centrifuge User Manual
Page 7

3
Symbols and Definitions (cont.)
IVD
For in vitro
diagnostic use
Clarifies for use as in vitro diagnostic only
Non sterile
Indicates non-sterile product
S N
Serial Number
Indicates instrument serial number code
Consult
Instructions
Consult instruction manual or insert sheet for further
explanation
Biological Risk
Universal precautions should be followed on all
specimens
Error Indicators
Error codes for the StatSpin MP are specified by a combination of the Error/Service LED flashing or
continuous and any combination of the cycle LEDs on the front panel.
Error/Service =
Error/Service light will continuously illuminate when the centrifuge has achieved a total cycle count of 18,000, which
is the useful life of the drive system. Drive mechanism needs replacing. Contact an authorized service center.
Error/Service
light flashing
or continuous
Rotor failed
to reach rpm
within 30 s
Cover opened
or cycle
changed during
operation
Insufficient
power to
maintain
rpm
Rotor spun
backward
direction during
reverse braking