Iris Sample Processing Statspin® MP Multipurpose Centrifuge User Manual
Page 11

7
Cycle Selection
Table 1: StatSpin MP Cycle Settings. The operator should experiment with different settings to achieve optimum
performance for specific applications. The following are general guidelines:
Setting
RPM/RCF
Time
Rotor
Urine
9,800/3,900
45 seconds
RT12
Hematocrit
16,000/13,700
120 seconds
RH12
Tube Rotor
15, 800/12,000
30 seconds
RT12/RM02
PlasmaRotor
®
20,000/11,200
30 seconds
RD01
LipoClear
®
15,800/12,000
95 seconds
RT12
Coag Plasma
20,000/11,200
120 seconds
RD01
Instructions for Use
1. Lift cover and install rotor.
2. Close and latch cover.
3. Choose desired cycle by depressing the “SET” button until the appropriate LED is illuminated.
4. Press
start button.
5. Upon completion of the cycle, the rotor decelerates to a complete stop in 10 seconds and the latch
interlock automatically unlocks.
6. Squeeze the black latch pieces together to open cover.