Appendix – Iris Sample Processing Statspin® MP Multipurpose Centrifuge User Manual
Page 17
13
Appendix
Appendix A - Specifications
Product No.
SSMP (supplied with 2 rotors, RT12 and RH12)
SSMP-1 (supplied with 1 rotor, RT12)
SSMP-2 (supplied with sample of PlasmaRotors)
Model No.
M901
Cycles/Speeds
Urine 9,800 (3,900 x g); 45 seconds
Hematocrit 16,000 (13,700 x g); 120 seconds
Tube Rotor 15,800 (12,000 x g); 30 seconds
PlasmaRotor
®
20,000 (11,200 x g); 30 seconds
LipoClear
®
15,800 (12,000 x g); 95 seconds
Coagplasma 20,000 (11,200 x g); 120 seconds
Acceleration Time
Approximately 6 seconds
Deceleration Time
Approximately 10 seconds
Electrical
24 Volts, DC, 1.7 amp. Includes switching power supply for 100-240
VAC, 50/60 Hz
Dimensions
Diameter 6.6”/ 16.25 cm
Height 6.3”/13.2 cm
Weight 5.5 lbs/2.5 kg
Environmental
Indoor use
Altitude up to 2000m
Temperature 5ºC to 40ºC
Maximum relative humidity 80% for temperatures up to 31ºC decreasing
linearly to 50% relative humidity at 40ºC
Main supply voltage fluctuations not to exceed +/- 10% of the nominal
voltage
Transient over-voltages according to installation category II
Pollution degree 2