beautypg.com

Myron L 6P and 4P User Manual

Page 36

background image

32

XVII. CARE and MAINTENANCE
Ultrameter IIs should be rinsed with clean water after use. Solvents should

be avoided. Shock damage from a fall may cause instrument failure.

A. Temperature Extremes

Solutions in excess of 71°C/160°F should not be placed in the cell

cup area; this may cause damage. The pH sensor may fracture if the

Ultrameter II temperature is allowed to go below 0°C/32°F. Care should

be exercised not to exceed rated operating temperature.

Leaving the Ultrameter II in a vehicle or storage shed on a hot day

can easily subject the instrument to over 66°C/150°F. This will void the

warranty.

B. Battery Replacement

Dry Instrument THOROUGHLy. Remove the four (4) bottom screws.

Open instrument carefully. Carefully detach battery from circuit board.

Replace with 9 volt alkaline battery. Replace bottom, ensuring the

sealing gasket is installed in the groove of the top half of case. Re-install

screws, tighten evenly and securely.

NOTE: Because of nonvolatile EEPROM circuitry, all data stored in

memory and all calibration settings are protected even during power

loss or battery replacement. However, loss of time and date may occur if

battery is removed for longer than 3 minutes (180 seconds).

C. pH/ORP Sensor Replacement (6P)

Order model RPR. When ordering, be sure to include the model and

serial number of your instrument to ensure receipt of the proper type.

Complete installation instructions are provided with each replacement

sensor.

D. Cleaning Sensors

1. Conductivity/TDS/Resistivity

The conductivity cell cup should be kept as clean as possible. Flushing

with clean water following use will prevent buildup on electrodes.

However, if very dirty samples — particularly scaling types — are

allowed to dry in the cell cup, a film will form. This film reduces accuracy.

When there are visible films of oil, dirt, or scale in the cell cup or on the

electrodes, use isopropyl alcohol or a foaming non-abrasive household

cleaner. Rinse out the cleaner and your Ultrameter II is again ready for

accurate measurements.

2. pH/ORP (6P)

The unique pH/ORP sensor in your Ultrameter II is a nonrefillable

combination type that features a porous liquid junction.

It should not be