Myron L 6P and 4P User Manual
Page 18
14
VII. CALIBRATION
A. Calibration Intervals
Generally, calibration is recommended about once per month with
Conductivity or TDS solutions. Calibration with pH solutions should
be checked twice a month. Calibration of ORP is not necessary (ref.
CALIBRATION INTERVALS, pg. 19).
B. Rules for Calibration of the Ultrameter II
1. Calibration Steps
a. Starting Calibration
Calibration is begun by pressing
while measuring Conductivity,
TDS or pH. Measuring continues, but the CAL icon is on, indicating
calibration is now changeable.
The reading is changed with the
and
keys to match the
known value. The calibration for each of the 4 solution types may be
performed in either conductivity or TDS mode.
Depending on what is being calibrated, there may be 1, 2 or 3 steps to
the calibration procedures.
Once in “CAL” mode, the
key becomes an “ACCEPT” key. At
each point, pressing
accepts the new calibration value and steps
you to the next adjustment (or out of CAL mode if there are no more
steps).
To bypass a calibration step, simply press
to accept the present
value as is.
b. Leaving Calibration
Calibration is complete when the “CAL” icon goes out. Pressing any
KCl, NaCl or 442
User
Cond Gain only
Tempco, then Gain
Res
Done in conductivity
Done in conductivity or TDS
TDS Gain only
Tempco, Ratio, then Gain
pH
7, acid and/or base (6P)
ORP Zero set with pH 7 automatically (6P)