Concentrator troubleshooting guide – Invacare IRC5LXO2 User Manual
Page 19

19
PART
ONE
CABINET
WARNING
UNPLUG the concentrator when cleaning. To avoid
electrical shock, DO NOT remove cabinet.
1. Clean the cabinet with a mild household cleaner and non-abrasive
cloth or sponge.
HUMIDIFIER
NOTE: To clean the oxygen humidifier, follow the instructions provided by
the manufacturer. If none are provided, follow these steps:
1. Clean the humidifier every day.
2. Wash it in soapy water and rinse it with a solution of ten (10) parts
water and one (1) part vinegar.
3. Rinse thoroughly with hot water and refill with distilled water to
the level shown on the humidifier.
WARNING
DO NOT overfill Humidifier Bottle.
SOLUTION:
1. Insert plug into outlet.
2. Inspect house circuit
breakers or fuses. If problem
recurs, use a different outlet.
3. Call Service Provider.
1a. Remove and clean
cabinet filters.
1b. Move oxygen concentrator
at least three (3) inches
away from walls, draperies
or furniture.
2. Do not use extension
cords. Move to another
electrical outlet or circuit.
3. Call Service Provider.
CONCENTRATOR
TROUBLESHOOTING GUIDE
PROBABLE CAUSE:
Main Power Loss:
1. Power cord NOT
plugged In.
2. No power at outlet.
3. Tripped circuit breaker.
System Failure:
1. Unit overheating due
to blocked air intake.
2. Insufficient power at
outlet.
3. Internal repairs
required
SYMPTOM:
Alarm:
SHORT BEEPS,
LONG PAUSE
Concentrator
NOT operating,
power switch ON.
Beep......Beep......
Alarm:
CONTINUOUS
Concentrator
NOT operating,
Power Switch ON.
Beeeeeeeep....
MAINTENANCE/TROUBLESHOOTING