beautypg.com

Hoffman Media DRIPMASTER EDD-4C User Manual

Page 15

background image

15

Item

number

(Figure 6)

Name

Control or
Indicator

Function

12

Drip Rate

Toggle switches

Sets the drip rate for the pump-on mode.
Choose only one toggle switch, and push it
up to the desired dpm value.

13

Timing

Toggle switches

These switches are set according to the
section "Setting Normal (pump on) Mode Drip
Rates", starting on page 16.

14 Drip

rate

7-segment
Display

A two-digit display shows the drip rate count

15

12 ' '

Yellow LED

Flashes every 12 seconds for timing
purposes. You can count the number of drips
seen during this 12-second interval and
multiply them by five to arrive at the dpm rate.

This LED is useful when you want to quickly
stabilize the drip rate by pressing the Open
Valve (SW16) or Close Valve (SW18) push
button.

16

FR

Green LED

Lights to indicate the six-second freeze
interval. During this interval, which occurs
once per minute, the needle valve may be
opened or closed in order to adjust the drip
rate.

17

Display switch

Toggle switch

Controls the two-digit numerical display:

• In the down position (recommended):

When pump is on, displays the last minute
drip count for a period of six seconds. The
display goes off until the next minute drip
count is displayed. When the pump is off,
the display remains off.

• In the up position (used for factory tests

only).