Emperature, Afety, Ontroller – HP 11744195 User Manual
Page 11
Chart Environmental Chambers
Features
the air-flow plenum and is not directly visible. The following figure is a picture of the
Watlow controller:
Watlow Series 96
temperature
controller.
❏
Temperature Safety Controller
A temperature Hi/Lo limit control protects the chamber against thermal runaway due to a
primary controller fault or control device failure. The input to the Hi/Lo limit control is a
type T thermocouple. The thermocouple is mounted in the air-flow plenum and is not
directly visible. The temperature safety controller is located inside the control box and is
not to be adjusted by the operator.
6