Honeywell 9782 Series User Manual
Page 41

Specifications and Model Number
7/99
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual
2-9
Note 1:
Customer must supply tagging information; Up to 3 lines allowed.
(22 characters for each line)
Note 2:
This option for use with HPW7000 assembly only.
Note 3:
Standard range should be selected if measuring the following
conductivity, resistivity or TDS ranges:
STANDARD RANGE
Conductivity
TDS
Cell Constant
0-1.999
0-1999 ppb
0.01
0-19.99
0-19,990 ppb
0.1
0-199.9
0-199.9 ppm
1.0
0-1999
0-1999 ppm
10
0-19,990
0-10,000 ppm
25 or 50
Resistivity
Cell Constant
0-19.99M
0.01
0-1.999M
0.1
0-199.9k
1.0
0-19.99k
10
0-1.999k
25 or 50
Note 4:
Wide range should be selected if measuring the following
conductivity, TDS or weight % concentration ranges:
WIDE RANGE
Conductivity
TDS
Cell Constant
100-1999
0-1999 ppm
0.1
1000-19,990
0-19,990 ppm
1.0
10-199.9 mS
0-199.9 ppt
10
10-500 mS
25
20-1000 mS
50
Weight % Concentration
Cell Constant
0-18.00% Hydrochloric Acid
50
0-19.99% Sulfuric Acid
50
0-19.99% Sodium Hydroxide
25 or 50
0-19.99% Sodium Chloride
25 or 50