Electrical specification, Auto power off, Battery changing – Amprobe AMB-45 Digital-Megohmmeter User Manual

Page 9: Fuse changing

background image

5

ELECTRICAL SPECIFICATION:

AC Voltage

Range

Resolution

Accuracy

600V

0.1V

1.5%rdg+5dgts

DC Voltage

Range

Resolution

Accuracy

600V

0.1V

1%rdg+3dgts

Ω Ohms (Autoranging)

Range

Resolution

Accuracy

400Ω

0.1Ω

1%rdg+5dgts

4000Ω

Continuity Beeper

Range

Active

Protection

40Ω

250Vrms

MΩ ( Autoranging )

Range

Resolution

Accuracy

4/40/400/4000MΩ ( 250V)

1KΩ

3%rdg+5dgts < 2GΩ
5%rdg+5dgts < 4GΩ

4/40/400/4000MΩ ( 500V)

4/40/400/4000MΩ ( 1000V)

Recording Length: 4000 records

AUTO POWER OFF

When the tester is idle for thirty minutes, with no function selector or button operation, it will turn itself off
automatically. To turn the tester on again, the user has to turn the function selector to “OFF” position, then the
selected function.

BATTERY CHANGING

When

appears on the LCD, the battery need to be replaced with new ones. To replace the battery the

user should turn the function selector to OFF position. Then the user needs to open the battery compartment
cover with a screwdriver. Eight AA 1.5V batteries are needed to replace the old ones. After all Batteries are
changed, put the cover back and fasten the screw.

FUSE CHANGING

When connect the meter under Ohm rang to a source > 10V, the protection fuse will break the circuit and new
one will be needed for replacement. To replace the fuse, user should prepare an F 0.5A 600V fuse. First, the user
should turn the tester off and remove the test leads then remove the Back Cover and Replace the fuse.