SAFETY INFORMATION

• Read the following safety information carefully before attempting to operate or service the meter.
• The circuit under test must be de-energized and isolated before connections are made except for voltage

measurement.

• Circuit connections must not be touched during a test.
• After insulation test, capacitance circuits must be allowed to discharge before disconnecting the test leads.
• To avoid damages to the instrument do not apply the signals, which exceed the maximum limits shown in

the technical specification tables, to the meter.

• Do not use the meter or test leads if they look damaged. Use extreme caution when working around bare

conductors or bus bars.

• Use the meter only as specified in this manual; otherwise, the protection provided by the meter may be

impaired.

• Caution when working with voltages above 60V DC or 30 V AC RMS. Such voltages pose a shock hazard.
• Before taking resistance measurements or testing acoustic continuity, disconnect circuit from main power

supply and all loads from the circuit.

ENVIRONMENT CONDITIONS:
• Installation Categories III 600V, Categories II-1000V
• Pollution Degree 2
• Altitude up to 2000 meters
• Indoor use only
• Relatively humidity 80% max.
• Operation Temperature 0 ~ 40

MAINTENANCE & CLEANING:
• Only qualified personnel should perform repairs or servicing not covered in this manual.
• Periodically wipe the case with a dry cloth. Do not use abrasives or solvents on these instruments.

GENERAL SPECIFICATION:

Display: 70 x 46mm Big LCD Panel with 60 segments analog bar Indication.
Over range Indication: “OL” will be shown on the LCD Panel when out of range measurement is made,

except for insulation resistance test, where the “HI“ will be shown.

Low Battery Indication: The will be show when the battery need to be changed.
Sampling Rate: 2.5 times/sec Digital Display; 10 times/sec Bar Graph
Power Source: 1.5V AA size Battery X 8
Operating Temperature and Humidity: 0

℃ to 40℃ (32℉ to 104℉), below 80% RH

Storage Temperature: -10

℃ to 60℃ (14℉ to 140℉)

Dimension: 230 x 116 x 62 mm; 9 x 4.6 x 2.4 inch
Weight: Approx. 750g (with battery)
Accessories: Test Lead, Large Jaw Alligator Clips, Battery, Carrying Case, Software, RS-232 Cable, Users

Manual.