Measuring functions – Amprobe AMB-45 Digital-Megohmmeter User Manual

Page 10

background image

6

MEASURING FUNCTIONS

ACV FUNCTION:

Turn function selector to voltage range. Connect black test lead to Black terminal and red one to
the Red terminal. Connect test lead to the test circuit in parallel. Operator can hold the reading by
pressing the

key.

DCV FUNCTION:

Turn function selector to voltage range. Connect black test lead to Black terminal and red one to the Red
terminal.

Press the

button to change the mode form ACV to DCV function. Connect test lead to the test circuit in

parallel. One can hold the reading by pressing the

key.

LOW OHM FUNCTION AND CONTINUITY FUNCTION:

Turn function selector to Ohm range. Connect black test lead to Black terminal and red one to the Red terminal.
Connect test lead to the test circuit in parallel. If the reading is less than 40 Ohm, the continuity beeper will
sound.

One may null the lead resistance (under 40 Ohm) by shorting the test lead than press the

key. When

the lead resistance is recorded a ZERO symbol will display on the LCD. Press

key again to go back to

normal operation. If the lead resistance is great than 40Ω, the error beeping will sound. Because the test current
provided by the meter could reach 200mA, do not use this range to test electronic component like diode,
transistor or fuse.

Caution:

Before measuring, verify the circuit is not live by voltage function

MEGAOHM FUNCTION:

Turn the function selector to the desired test voltage range. The LCD will display “----” to indicate the tester is
standing by. Connect black test lead to Black terminal and red one to the Red terminal. Connect test lead to the
test circuit in parallel. One may take the measuring under manual power mode or power lock mode: