12
Specifications
T/C Type
Range Resolution
Accuracy
K
-200 to 650
°C 0.1
°C
651 to 1370
°C 1.0
°C
(0.1% rdg + 0.7
°C)
-328 to 1000
°F 0.1
°F
1001 to 2498
°F 1.0
°F
(0.1% rdg + 1.4
°F)
J
-200 to 500
°C 0.1
°C
501 to 760
°C 1.0
°C
(0.1% rdg + 0.7
°C)
-328 to 940
°F 0.1
°C
941 to 1400
°F 1.0
°F
(0.1% rdg + 1.4
°F
T
-200 to 390
°C 0.1
°C
(0.1% rdg + 0.7
°C)
-328 to 730
°F 0.1
°F
(0.1% rdg + 1.4
°F)
R/S
0 to 1000
°C 0.1
°C
1001 to 1760
°C
1.0
°C
(0.3% rdg + 0.7
°C)
32 to 1000
°F 0.1
°F
1001 to 3200
°F 1.0
°F
(0.1% rdg + 1.4
°F)
E
-200 to 380
°C 0.1
°C
381 to 736
°C 1.0
°C
(0.3% rdg + 0.7
°C)
-328 to 720
°F 0.1
°F
721 to 1832
°F 1.0°F
(0.1% rdg + 1.4
°F)
Additional Specifications
OPERATING CONDITIONS
0 to 50
°C (32 to 122 °F) at 0 to 85 % relative humidity
(non-condensing)
Ambient Coefficient
0 to 18
°C and 28 to 50 °C (Ambient temperatures) For each °C
ambient below 18
°C or above 28 °C, add the following tolerance into
the accuracy spec: 0.01% of reading +0.03
°C (0.01% of reading
+0.06
°F).
TMD90A_all.pdf 16
3/23/2007 3:22:50 PM
- AC71B Clamp-On-Multimeter (96 pages)
- AC50A (78 pages)
- ACD-10-TRMS-Pro (16 pages)
- AC68C Clamp-Multimeter (52 pages)
- ACD-14-TRMS-FX Clamp-On-Multimeter (19 pages)
- ACD-10-TRMS-PLUS ACD-10-PLUS Clamp-Multimeters (116 pages)
- ACD-40PQ Clamp-Meters (16 pages)
- ACD-23SW Digital-Clamp-Meters (20 pages)
- ACD-41PQ Clamp-On-Power-Meters (20 pages)
- ACD-41PQ Clamp-On-Power-Meters (121 pages)
- ACD-14-TRMS-PLUS Clamp-On-Multimeters (148 pages)
- ACD-330T Clamp-On-Multimeter (42 pages)
- ACD-21SWC Digital-Clamp-Meters (84 pages)
- ACD-4 Mini-Clamp-DMM (16 pages)
- ACD-6-TRMS-PRO Clamp-On-Multimeters (15 pages)
- FLASH-80 Industrial-Mini-Flashlight (64 pages)
- FLASH-80 Industrial-Mini-Flashlight (46 pages)
- FLASH-80 Industrial-Mini-Flashlight (44 pages)
- FLASH-80 Industrial-Mini-Flashlight (98 pages)
- ACD-50NAV ACD-51NAV ACDC-52NAV ACD-53NAV ACD-54NAV Navigator-Clamps (36 pages)
- ACD-55HPQ Clamp-Meter (62 pages)
- ACDC-620T Clamp-On-Multimeter (65 pages)
- ACDC-400 Digital-Clamp-On-Multimeter (98 pages)
- ACDC-100-TRMS Clamp-On-Multimeters (97 pages)
- RS-3 Rotary-Scale-Clamp-On (4 pages)
- AD105A Clamp-Meter (34 pages)
- LH41A Clamp-On-Ammeter (53 pages)
- 30XR-A Professional-Digital-Multimeter (53 pages)
- 33XR-A Professional-Digital-Multimeter (73 pages)
- 34XR-A Professional-Digital-Multimeter (73 pages)
- 37XR-A Professional-Digital-Multimeter (86 pages)
- 5XP-A 15XP-A 35XP-A Compact-Digital-Multimeters (126 pages)
- AM-110-TRMS Digital-Multimeters (15 pages)
- AM-160-A DMM (30 pages)
- 38XR-A Professional-Digital-Multimeter (92 pages)
- AM-33 Digital-Multimeter (1 page)
- AM-34 Auto-Digital-Multimeter (1 page)
- AM-47 Multimeters (8 pages)
- AM-240 Digital-Multimeters (75 pages)
- AM-250 Digital-Multimeter (22 pages)
- AM-270 Industrial-Multimeter (28 pages)
- AM-60 Compact-Digital-Multimeter (20 pages)
- AM8C Analog-Multimeter (66 pages)
- CR50A Capacitance-Resistance-Meter (46 pages)
- AM91RS Digital-Multimeter (30 pages)