Amprobe SOLAR-600 Solar-Analyzer User Manual
Page 18

11
Auto Scan
1. Press
button to turn on the Analyzer.
2. Properly connect the pair of Kelvin clips to the solar panel and the
Analyzer. The red Kelvin clip is for positive pole and the black Kelvin clip
for negative pole. (refer to above Connecting Diagram).
3. Turn on any available light source (e.g. halogen lamp, xenon lamp,
tungsten lamp, …) and let it illuminates solar panel uniformly.
Or place the solar panel under the sun.
4. Press
(AUTO SCAN) button to perform Auto-scan. After the
scanning is finished, the result will come out like below.
5. The unit automatically measures the followings parameters: Vopen,
Ishort, Pmax, Vmaxp, and Imaxp. Based upon those parameters, the unit
run simulation and draw I-V / V-I curve and P-V / P-I curves in LCD.
6. Users can move the cursor to review each individual values along the
curve.
�
WARNING!
There is a time delay before the unit performs the “Auto Scan”. This time
delay allows the light source to be turned on before “Auto Scan” starts.
Time delay can be set in the SETUP menu
Note: If the short circuit current (Ishort) exceeds 12A, Auto Scan will not be
performed. Please select Manual Scan and limit the ending value of Scan
to be less than 12A.