Instruction Manual

748297-E
June 2003

Description and Specifications

Rosemount Analytical Inc. A Division of Emerson Process Management

1-4

Model NGA2000 HFID

1-5 SPECIFICATIONS

a. General

Measurement Species ....................Total hydrocarbons

Ranges (H

2

/He Fuel)

Low range ............................0 to 10 ppm, CH

4

, through 0 to 1%, CH

4

at an oven setpoint between

113

°C and 191°C

High range ...........................0 to 50 ppm, CH

4

, through 0 to <5%, CH

4

at an oven setpoint

between 113

°C and 191°C

Analysis Temperature .....................Adjustable from 200

o

F to 400

o

F (93

o

C to 204

o

C), maintained within

±11

o

F (±6

o

C) from the setpoint.

Repeatability ....................................

≤1% of fullscale for successive identical samples, at a constant
temperature, sample flow and fuel, burner air, regulated air and
sample pressures

Min. Detectable Level......................0.10 ppm, CH

4

Noise ................................................

≤1% of fullscale, peak to peak

Linearity............................................

≤±1% of fullscale, ≤±2% of data point (must be above the minimum
detectable level)

Response Time ...............................

≤1.5 sec., 0% to 90% of fullscale

Drift

Zero .....................................

≤ ±1% of fullscale/24 hours at constant temperature, sample flow,

hydrocarbon concentration of supply gases, and fuel, burner air,
regulated air and sample pressures.

Span ....................................

≤ ±1% of fullscale/24 hours at constant temperature, sample flow,

hydrocarbon concentration of supply gases, and fuel, burner air,
regulated air and sample pressures.

Effect of Temperature....................

≤±2% of fullscale for any ambient temperature change of 10

o

C and

rate of change less than 10

o

C/hr.

Operating Temperature ................. 59

°F to 95°F (15°C to 35°C)

Power Requirements.......................+24 VDC ±5%, 120 W max. direct to Analyzer Module

Ripple and Noise .................<100 mV pp
Line and Load Regulations..<1%