RIGOL DG5000 Series User Manual

Page 58

background image

RIGOL

Quick Guide for DG5000

12

9. CH2 ExtTrig In (ExtTrig2)

This SMB connector accepts an external TTL-compatible pulse as the trigger
input of CH2. Besides, it can also be used as the trigger out in Sweep and Burst
mode.

10. (11.)10MHz In/10MHz Out

These two connectors are used to synchronize generators. The connector
[10MHz In] accepts an external 10 MHz clock signal, and the connector [10MHz
Out] can output a 10 MHz clock signal generated by the crystal inside the
generator.

19. LAN

Through this interface, the generator can be connected to your local network for
remote control. An integrated testing system may be built, as the generator
conforms to the LXI-C class standard of LAN-based instrument control.

20. USB Device

Through this interface, the generator can be connected to a PictBridge printer to
print its screen, or be connected to a PC and controlled via PC software.

21. GPIB

Meet IEEE-488.2 specification.

22. USB Host

Reference to USB Host page 5.

23. Lock Hole

Use the security lock (please buy it yourself) to lock the generator in fixed
location.

24. Power Switch

Connect or cut off the power supply.

25. Power Socket

The generator can accept two types of AC power supply.
AC Power Supply: 45-440 Hz, 100-127 V, or 45-60 Hz, 100-240 V.
Power Fuse: 250 V, T3A.
Power Consumption: less than 125 W.