Warning, Caution – Precision Medical EasyDial Reg Oxygen Regulator User Manual

Page 6

background image

10. To remove Dial Regulator from cylinder:

• Completely close oxygen cylinder valve.
• Turn dial until the Indicating Pointer is pointing to “OFF”.
• S-L-O-W-L-Y loosen T-handle or Nut and remove Regulator.

wARNING

To prevent risk of burns, fire or injury to person(s):
• Prior to handling Dial Regulator and cylinder valve, be sure hands are

free of oil, grease and other contaminates.

• DO NOT use Regulator if oil, grease or dirt is found.

Particles propelled in a high-velocity oxygen stream could create a
spark, causing a fire. Oil and grease can react with oxygen and other
oxidizers, causing an explosion.

• There is NO FLOw between settings. The Indicating Pointer

must point to a specific number on the dial to obtain flow.

CAUTION

DO NOT use tools to tighten T-handle. This can lead to over tightening
and will cause damage to the Dial Regulator.

CAUTION

• Inspect the Dial Regulator for visual damage before use, DO NOT USE

if damaged.

• For optimum safety, use only a Precision Medical, Inc. Yoke Seal part

number 1221. The Yoke Seal supplied with the Dial Regulator is reusable.

• DO NOT use plastic yoke seals.

Tell us how we are doing!

Visit us at www.precisionmedical.com