Warning, Caution – Precision Medical EasyDial Reg Oxygen Regulator User Manual
Page 6
10. To remove Dial Regulator from cylinder:
• Completely close oxygen cylinder valve.
• Turn dial until the Indicating Pointer is pointing to “OFF”.
• S-L-O-W-L-Y loosen T-handle or Nut and remove Regulator.
wARNING
To prevent risk of burns, fire or injury to person(s):
• Prior to handling Dial Regulator and cylinder valve, be sure hands are
free of oil, grease and other contaminates.
• DO NOT use Regulator if oil, grease or dirt is found.
Particles propelled in a high-velocity oxygen stream could create a
spark, causing a fire. Oil and grease can react with oxygen and other
oxidizers, causing an explosion.
• There is NO FLOw between settings. The Indicating Pointer
must point to a specific number on the dial to obtain flow.
CAUTION
DO NOT use tools to tighten T-handle. This can lead to over tightening
and will cause damage to the Dial Regulator.
CAUTION
• Inspect the Dial Regulator for visual damage before use, DO NOT USE
if damaged.
• For optimum safety, use only a Precision Medical, Inc. Yoke Seal part
number 1221. The Yoke Seal supplied with the Dial Regulator is reusable.
• DO NOT use plastic yoke seals.