7

CLEANING / MAINTENANCE

WARNING

• Turn the Easy Air Compressor off and disconnect from any

external power source before cleaning.

DO NOT attempt to repair. There are no serviceable parts inside.

To clean the filter (weekly):

1. Unplug Compressor.
2. Check air inlet filters for dust buildup.
3. If dust buildup occurs, remove air inlet filters and wash in warm, soapy

water and rinse.

4. Absorb excess water from filter with a dry towel.
5. Return the air inlet filters.

Cleaning for Exterior Components:

Use a clean cloth or sponge dampened with soap and water. Wipe dry

with a clean cloth or paper towel.

CAUTION

• Excessive dust buildup on filter will reduce performance of

Compressor. If this occurs, clean or replace with a new filter.

The air filter does not filter out saline. Treatments using nebulized

saline will draw the saline mist back into the Compressor. Saline

causes rapid deterioration of the compressor seals and will void the

warranty.

At ~8500 Hours

(or as needed)

:

Return the Easy Air Compressor to your Provider for Compressor

Maintenance (including rebuilding pump).

Contact Precision Medical, Inc. for Preventative Maintenance Services.