TROUBLESHOOTING

If the Easy Air Compressor fails to function, consult the Troubleshooting

Guide below. If problem cannot be solved, consult your Provider.

Problem

Probable Cause

Remedy

Compressor

does not

operate

1. On/Off Switch in

“OFF” position

2. Power Cord not

connected properly

3. Compressor

overheated

4. Blown Fuse
5. Power Interruption

6. Trapped back

pressure

1. Push On/Off Switch to

“ON” position

2. Check Power Cord

connections

3. Shut Compressor

Off, Wait 15 minutes,

Ensure vents are clear.

4. Replace Fuse
5. Turn off Compressor.

When Power restored,

repeat Start Up

Procedure.

6. BEFORE STARTING

COMPRESSOR.

Pull “OUT” regulator

Knob to unlock,

then turn Knob fully

counterclockwise.

Unable to adjust

pressure / flow

1. Locked Regulator

Knob

1. Pull Regulator Knob

out

Overheating

1. Poor Ventilation

2. Dirty Air Filters

1. Allow a minimum of

4 in (10 cm) clearance

on all sides to ensure

adequate air flow

for cooling of the

Compressor.

2. Clean Air Filters

Low pressure /

flow

1. Loose connection at

outlet fitting

2. Improper regulator

adjustment

1. Tighten connection

2. Re-adjust regulator