TROUBLESHOOTING
If the Easy Air Compressor fails to function, consult the Troubleshooting
Guide below. If problem cannot be solved, consult your Provider.
Problem
Probable Cause
Remedy
Compressor
does not
operate
1. On/Off Switch in
“OFF” position
2. Power Cord not
connected properly
3. Compressor
overheated
4. Blown Fuse
5. Power Interruption
6. Trapped back
pressure
1. Push On/Off Switch to
“ON” position
2. Check Power Cord
connections
3. Shut Compressor
Off, Wait 15 minutes,
Ensure vents are clear.
4. Replace Fuse
5. Turn off Compressor.
When Power restored,
repeat Start Up
Procedure.
6. BEFORE STARTING
COMPRESSOR.
Pull “OUT” regulator
Knob to unlock,
then turn Knob fully
counterclockwise.
Unable to adjust
pressure / flow
1. Locked Regulator
Knob
1. Pull Regulator Knob
out
Overheating
1. Poor Ventilation
2. Dirty Air Filters
1. Allow a minimum of
4 in (10 cm) clearance
on all sides to ensure
adequate air flow
for cooling of the
Compressor.
2. Clean Air Filters
Low pressure /
flow
1. Loose connection at
outlet fitting
2. Improper regulator
adjustment
1. Tighten connection
2. Re-adjust regulator