GROUNDING INSTRUCTIONS

The Easy Air Compressor must be plugged into an electrical outlet that is

properly grounded. In the event of an electrical short circuit, grounding reduces

the risk of electrical shock. This product is equipped with a grounding circuit.

DANGER

Improper use of the grounding plug can result in a risk of electrical

shock.

WARNING

Ground reliability can only be achieved when the power cord is

connected to an equivalent receptacle marked “Hospital Grade”.

OPERATING INSTRUCTIONS

CAUTION

Inspect the Easy Air Compressor for visible damage, DO NOT USE

if damaged.

1. Become familiar with the Compressor (see Component Identification).
2. Position the Compressor near a properly grounded electrical

receptacle. Allow a minimum of 4 in (10 cm) clearance on all sides to

ensure adequate air flow for cooling of the Compressor.

3. Ensure the On/Off (I/O) Switch is in the “OFF” (O) position.
4. Ensure the Power Cord is properly connected.
5. Attach appropriate device to the Outlet Fitting securely, without

overtightening.

6.

Pull Regulator Knob “OUT” to unlock regulator, then turn Knob

fully counterclockwise. (Failure to do so may cause the device

not to start.)

7. Press the On/Off (I/O) Switch to the “ON” (I) position.
8. Adjust pressure/flow.
To

INCREASE pressure / flow - Turn Knob CLOCKWISE

To DECREASE pressure / flow - Turn Knob COUNTERCLOCKWISE

9. Push Regulator Knob “IN” to lock pressure / flow setting.

10. Compressor is now ready to use.
11. When the Compressor is not in use, press the On/Off (I/O) Switch to

the “OFF” (O) position.