Figure 6-9. Fault 5, O

2

Cell Open

e. Fault 5, O

2

Cell Open

The O

2

cell connection is open. The fault

displays as shown in Figure 6-9.

1. Allow adequate time for the oxygen

probe to reach operating temperatures.
Probe temperatures below approxi-
mately 500

°C (932°F) may result in this

fault.

2. Refer to Figure 6-4 and check the O

2

cell wiring connections at terminals 7
and 8. Ensure the wires are properly
connected.

NOTE

Check the cell output voltage at the
probe terminals -- not at the
electronics.

3. Apply low calibration check gas (0.4%

O

2

). Measure the cell output from the

O

2

cell wires at the probe terminal

block. The cell output should be 100
±20 mV. If no voltage can be meas-
ured, the cell is open.

4. If the O

2

cell is open, replace the oxy-

gen probe per paragraph 5-2.

%

26020029

Figure 6-10. Fault 6, Cell Impedance Too High

f. Fault 6, Cell Impedance Too High

The O

2

cell impedance has exceeded 100

ohms. The fault displays as shown in Figure
6-10.

1. This fault is usually indicated in con-

junction with Fault 5, Cell Open. Cor-
recting Fault 5 should correct Fault 6.

2. If Fault 6 appears independently, the

cell has degraded beyond
specification.

3. If the O

2

cell has become too old, re-

place the oxygen probe per paragraph
5-2.