Instruction Manual

IB-106-3081 Rev. 1.5
September 2002

3-10 Startup and Operation

Rosemount Analytical Inc. A Division of Emerson Process Management

Model 3081FG

%

O CONCENTRATION

2

TRACK O

DURING CALCHECK

2

26020051

Figure 3-18. SET O

2

TRACKING

l. SET

O

2

TRACKING

Refer to Figure 3-18. Use this screen to
permit the 4-20 mA line to track the O

2

value

during a calibration check. Press ENTER to
begin editing. Use the arrow keys to select
Y or N. Entering Y (yes) will allow the 4-20
mA line to track the O

2

value during the

calibration check. Entering N (no) will hold
the O

2

value steady during the calibration

check. Press ENTER to accept the value.
Press NEXT to display the SET CODE
screen.

m. SET CODE

Refer to Figure 3-19. Use this screen to set
the security code for the Model 3081
Transmitter. Press ENTER to begin editing.
Use the arrow keys to select and change
the value. Select any value between 000
and 999, excluding 000 and 555. Code 000
indicates that no code is set. Code 555 ac-
cesses the DISPLAY CODE screen. Press
ENTER to accept the value. Pressing NEXT

%

O CONCENTRATION

2

26020016

Figure 3-19. SET CODE

returns to the FAULT VAL screen at the be-
ginning of the PROGRAM MENU.

n. Model 3081 Transmitter Parameters

Table 3-1 lists the range and default value
of operator-adjustable variables used by the
Model 3081 Transmitter. These variables
may all be changed from the PROGRAM
MENU screens.

Table 3-1. Model 3081 Transmitter Parameters

Parameter

Range

Default

Indicator Fault
Value

high value (24 mA)
low value (3.8 mA)

low value
(3.8 mA)

Upper Range Limit

2.0-25.0%

10.0%

Upper Cell Temp.
Fault Value

650-1600

°C

1600

°C

Analyzer Access
Code

000-999 (excluding
000 and 555)

000 (no
code)