Resuming pump delivery, Resume – Energizer MMT-511 User Manual
Page 91
Suspend 79
4. After 30 seconds, your pump will return to the HOME screen
which now will show solid circles, indicating your pump is in
the Attention mode.
5. Press ESC to show the STATUS of your suspended pump.
Resuming pump delivery
1. Press ACT from the HOME screen and the RESUME screen will
appear.
2. Press ACT again to restart basal insulin delivery. Pump will
beep once.
3. The pump will return to the HOME screen and will resume
your previously programmed basal insulin deliveries.
MiniMed
MiniMed
STATUS 11:30A U100u
Suspended at 11:20A
Last Bolus N 3.8U
7:58A 22 AUG
pump
suspended
MiniMed
Stopped at 11:20A
RESUME
ACT-Restart Basal
u