Easy bolus, Infusion pressure, Insulin concentration/type – Energizer MMT-511 User Manual
Page 157: Normal bolus

Pump Specifications 145
Easy bolus
■
Features programming using audible tones (or vibrate
pulses) in user-determined increments.
Beep mode range: 0 to maximum bolus
Vibrate mode range: 0 to 20 steps or maximum bolus,
whichever comes first.
■
Step size < maximum bolus.
■
Adjustable step size: 0.1 to 2.0 units per step.
■
Accessible from Remote Control or pump buttons.
Infusion pressure
■
Maximum infusion pressure and occlusion pressure is
13.7 PSI.
Insulin concentration/type
■
Pump users can select user-filled U50 and U100.
Normal bolus
■
Range 0.1–25.0 units of insulin (limited by maximum bolus
setting).
■
The bolus delivery rate is fixed by concentrations as follows:
■
U100: up to 1.5 units/minute
■
U50: up to 0.75 units/minute